Instrument using
measured DAR:
Select more than 1 minute length measurement.
Select the appropriate voltage and start.
The instrument automatically record 15 s and 60 s resistance, and calculate the absorption than after 60 seconds.
Absorption ratio calculation method: DAR = R60 / R15;
measured PI :
Select more than 10 minute length measurement.
Select the appropriate voltage and start.
The instrument automatically record 15S, 60S and 600S resistance, and calculate the polarization index after 600 seconds.
Polarization index calculation method : PI= R600/R60;
test connection points for attention
1. Tested to confirm product safety grounding, try not charged.
2. Confirm the instrument E end (ground) have been grounded.
3. The use of G terminal (protection ring) (this machine is of low voltage side of the screen)
When measuring high insulation resistance, the sample should be in the position a conductor surface protection between the two measuring end ring, and the conductor protection ring is connected to the instrument with a test line G side, try to eliminate the measuring error caused by the surface leakage current, security testing is accurate.link:https://www.beijing-huazheng.com/transformer-test-equipment/hzjy-20k-20000v-20kv-digital-insulation.html